Unity Lab Services™ Thermo Scientific™ Ultra-Low Temperature Freezers Validation Service

Stay compliant with our Industry-Exclusive, No-Charge Requalification Guarantee included with each Ultra-Low Temprature Freezers Models: Thermo Scientific TSX-series (post 2018), Forma 89000-series, HeraFreeze HLE, Revco RLE, Thermo Scientific TLE

Additional Details:
Additional Details: Weight: 0.00010kg



Product Code. 15671599

Quantity Price
1 £ 1726.0 / Each
Estimated Shipment
11-07-2025
Add to basket


  • Description and Specification
  • Accessories and Spare Parts
  • Special Offers
Description and Specification

Specification

For Use With (Equipment) Thermo Scientific Ultra-Low Temperature Freezers including Ultra-Low Temprature Freezers Models: Thermo Scientific TSX-series (post 2018), Forma 89000-series, HeraFreeze HLE, Revco RLE, Thermo Scientific TLE
Description Validation (IQ/OQ) includes a standardized protocol procedure, IQ: Verify Installation and environmental conditions, OQ: Verify operation to design specification, perform empty chamber temperature mapping, open door and power loss testing.
Location Contact your local sales representative for a list of standard and extended coverage areas.
Includes Travel to major metropolitan areas. Does not include Installation. Instrument must be installed, powered-up and set temperature point for 24 hours before on-site service

IQ

  • Unpacking, Assembly and Installation Verifies that the equipment, manuals, supplies and any other accessories arrived undamaged as specified in the sales order. Verifies equipment is assembled and installed and any initial diagnostics tests have been performed. Any abnormal event(s) observed during assembly and installation have been documented.
  • Utility, Facility and Environment Verifies that the installation site satisfactorily meets manufacturer-specified environmental requirements.

OQ System Component Information Detailed configuration information for each system component have been recorded.

  • OQ Limits: Record a list of manufacturer recommended limits for ensuring that the system is operating as expected.
  • Equipment Operational Tests and Results: Test important equipment functions to verify that the equipment operates as intended by the manufacturer and required by the user. This is a group of selected important equipment parameters for testing according to the intended use of the equipment.

- Key Tests: Door Ajar, Power Failure, Temperature Alerts

Accessories and Spare Parts
Special Offers